2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 442
TMP- 1.0.0 ‘BOB MOVING’ / ISOLATE POSSIBLE AIR LEAK
a) Small air leaks cause BIG problems! TWO (2) checks. Be THROUGH!
CHECK #1:
Using a flashlight, ENSURE no air bubbles moving into the machine through the acid and
bicarb inlet tubing!
CHECK #2:
ENSURE the external flow indicator’s (sight tube) fittings are tight!
b) Call debug screen 1. WITHOUT LOOKING AWAY, watch
FILACT
(middle column) for two (2) minutes OR until it EVER = 1, even
just once, indicates a Filling Program! TWO (2) possible
scenarios:
1)
IF (and ONLY if) FILACT EVER = 1:
Proceed to
page 410
, procedure number
2) IF FILACT ALWAYS = 0:
Call debug screen 4 to see
PDIA
(left
column) AND
ADIA
(right column). BOTH should be REMAINING
between 2.0 and 7.5. TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA AND ADIA are REMAINING between 2.0 and 7.5:
Proceed to
page 444
, procedure number
2) IF PDIA AND / OR ADIA is
IS NOT
remaining between 2.0 and 7.5:
See parts a
THROUGH e below:
a) Momentarily plug the acid concentrate into its rinse port to call “Select Program”!
b) Return the connector to acid concentrate!
c) Select the screen’s ‘Dialysis’ button but
DO NOT press ‘CONFIRM’ till
instructed!
d) Call debug screen 0. If parts a through c were performed correctly the TOP balancing
chamber valves, #31 through #34, ‘dots’ REMAIN BLUE!
e) Call debug screen 4 to watch
PDIA
(left column) AND
ADIA
(right column) for one (1)
minute. TWO (2) possible scenarios:
1)
IF (and ONLY if) PDIA AND ADIA remaining between 3.0 and 6.0 AND DOES
NOT change more than 0.1 i.e. stable:
2)
IF PDIA AND / OR ADIA is IS NOT between 3.0 and 6.0
OR one or both IS NOT stable:
See parts a AND b below:
a) Remove the RED DIALYZER connector from the
shunt door and
HOLD
it as seen in the Figure right!
Part b next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...