2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 471
TMP- 6.1.0 CREATE NEGATIVE PRESSURE
A)
The screen says
“6. Pressurize until dialysate pressure reads -250 mmHg….”
.
B)
PULL on the syringe plunger to achieve negative (-)250 +/- 5 mmHg on the external meter.
C)
Clamp the Four-Way’s syringe tubing segment
to
keep this pressure.
D)
ENSURING the external meter = -250 +/- 5 mmHg, is the screen’s
[Dial Pressure Reference]
window
between 130 and 150?
Yes
[Dial Pressure Reference]
is between 130 and 150! See part E.
No
[Dial Pressure Reference]
is NOT between 130 and 150! Proceed to
page 472
, procedure
E)
Press ‘CONFIRM’. Figure right, TWO (2) possible
scenarios based on if an “Operator Error” banner
occurs:
1) IF (and ONLY if) an “Operator Error” occurs:
(page 472).
2) IF an “Operator Error” did NOT occur:
See parts a THROUGH f below:
a) Press ‘CONFIRM’ twice to save the calibration then turn the machine off.
b) Turn the machine on and return to Dialysis Program (“Select Progr
am” → ‘Dialysis’ →
‘CONFIRM’)!
c) Call debug screen 1. If not already, allow
FILACT
= 0.
d) Press the Dialysate Flow on/off key. The Flow on/off lamp blinks indicating Flow is off!.
e) Call debug screen 0. ENSURE all eight balancing chamber valves (#31 through #38) are
REMAINING white (Dialysate Flow is OFF)! .
f) Return to
page 460
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...