2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 507
UF- 4.0.0 ISOLATE UF PUMP (#22) MECHANICS
a) Turn the machine OFF and disassemble the UF Pump.
b) Per the Figure right, TWO checks:
Check #1:
ENSURE a STRONG spring is
under the OUTPUT (colored)
nozzle AND its seal is pointing
INTO the pump; Replace the
spring if damaged or bent.
Check #2:
ENSURE a
WEAK spring is
under the INPUT (white)
nozzle AND its seal is pointing
AWAY from the pump. Replace
the spring if damaged or bent
c) Per the Figure below, FOUR (4) more checks:
Check #3:
ENSURE the diaphragm is NOT torn and for debris in the ports!
Check #4:
ENSURE the Retaining Clip is in the slot to hold the ‘Heavy Spring’ at its maximum
compression.
Check #5:
ENSURE a ‘shim washer’ has not fallen into the solenoid cavity which may ‘jam’ the
diaphragm shaft!
WARNING!
While checking keep track of ALL ‘shim washers’
Check #6
Check the solenoid and diaphragm assembly for corrosion.
d) Reassemble the pump and attach ALL tubing
HOWEVER,
DO NOT mount the pump to the cabinet yet!
page 509
Figure 88 – UF Pump Exploded
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...