2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 157
CLEAN- 1.0.5 ISOLATE VERR
NOTE:
From here forward, if (and ONLY if) a leak develops, address it first by proceeding to
page
a) Figure right, ENSURE a “No Water” alarm
*
NEVER
occurs!
*
From here forward, if (and ONLY if) a “No Water” alarm EVER occurs address it first!
b) Call debug screen 1 (Figure right).
Does
VERR
(lower right) = 0?
Yes
VERR
= 0! Proceed to
page 158
, procedure
No
VERR
= 1 OR more! See
procedure number
(page 157).
CLEAN- 1.0.6 VERR = 1 OR MORE
a) Call debug screen 0
(Figure right).
b) Ignoring a ‘blink to 1’ that
lasts less than two (2)
seconds, watch the
2
nd
window down,
Valve
Error
for one (1) minute!
c) Does
Valve Error
EVER = 1 for LONGER THAN two (2) seconds! TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve Error = 0 OR ‘blinks to 1’ for less than two (2) seconds!
ENSURING the
shunt door is CLOSED, proceed to
page 206
2)
IF Valve Error EVER = 1 LONGER THAN two (2) seconds:
Proceed to
page 189
, procedure
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...