![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 525](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635525.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 521
SECTION 16 - CHECKING THE ACTUATOR BOARD CABLE
The 50-conductor actuator cable runs between the distribution board and the Actuator-Test Board.
Check
ONLY the NOTED above component connection(s)!
a) Per the Figure below, this procedure checks the cable by measuring conductor
resistance (Ω) from one
end (distribution board) to the other (Actuator-Test Board). Good conductors measure 2
Ω
or less.
b)
IMPORTANT!
Turn the machine OFF to prevent damage! Open the card cage!
c) Check the cable’s entire length (distribution board to Actuator-Test Board) for bare wires or other damage!
d) Per the Figure above, unplug the cable from the distribution board’s ACTUATOR connector. Check inside
the vacant ACTUATOR connector for ‘white’ corrosion or damaged
male
pins.
Damage ANYWHERE
may be the problem!
Figure 91 – Female Distribution Board Connector
e) To determine your ohm
(Ω) meter’s internal resistance touch the leads together. The meter MUST read
less than 1
Ω
! Subtract the reading from ALL subsequent measurements!
f) Per the Figure above, the cable’s
female
connector has two rows of
PLUGS
. Hold it with its
KEY TAB
DOWN. This puts the even numbered
PLUGS
in the top row AND the ‘Rear of Machine’ to the right.
g) Refer to
(page 522), to LOCATE the appropriate numbered
PLUG(S)
for the NOTED PUMP
or
h) A meter probe will NOT penetrate deep enough and a
paper clip may
cause damage!
Use 25 gauge non-stranded wire as a ‘probe’. Per the
Figure (right), insert the ‘probe’ into the appropriate
PLUG
, then place
(or clip) one of the meter leads onto it.
i)
(page 522), the cable terminates at the actuator
board’s P2 connector. The rear side
PINS
are in two rows. Odd
numbered in the bottom row, even in the top. NOTING the reference to the ‘Rear of Machine’ locate the
appropriate
PIN
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...