T- 2.6.2 ISOLATE POWER CONTROL BOARD
A) Figure below, lay the power supply panel down to see the component side of the Power Control board.
Figure 48 – Power Control Board
B)
Per the Figure above
, TWO (2) checks:
CHECK #1:
At connector K1 and K2, gently yank on both wires to ENSURE they are securely attached!
CHECK #2:
Do resistors R10 and R11 show signs of burning?
Yes
Burning located! Replace BOTH the Power Control Board AND the Triac*. *To
LOCATE the Triac refer to the Figure below.
No burning! See procedure number
(page 280).