2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 384
CO- 10.0.36 ISOLATE POST COND CELL #13
a) From debug screen 5,
FPRE
changes continuously. RECORD its approximate
value for reference later!
b) Figure right, place Cond Cell’s #13 connector into
Conductivity Cell #7’s distribution board position
“X7, COND”.
c) Is
FPRE
/- 200 of what was recorded in
part a?
Yes
FPRE
/- 200 of what was
recorded! See procedure number
No
FPRE
is NOT /- 200 of what was
recorded. Either Cond Cell #13 is not in distribution board position “X7, COND” OR it* is bad.
*To LOCATE Cond Cell #13 refer to
Figure 62
(page 378).
CO- 10.0.38 VERIFY SENSOR BOARD
a) Return Cond Cell #13 to distribution board position “x13, COND-POS”.
b) Return Cond Cell #7 to distribution board position “x7, COND”.
c) Was a COND CELLS calibration performed in THIS troubleshooting session (Yes or No)?
Yes
Cond Cells calibration was already performed!
Read before performing!
Swap in a
known good Sensor Board then see (ABOVE) procedure number
page 354
).
If you return here swap in a known good Functional Board and calibrate Cond Cells.
No Cond Cells calibration was not performed! See (ABOVE) procedure number
page
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...