2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 582
HEATER RELAY- 2.0.2 HEATER RELAY TEST FAIL REOCCURED / RETEST HEATER CONTROL
a) Ensuring the machine is off, two (2) people are required for this procedure!
b)
IMPORTANT! Set your volt meter to AC voltage (~ V, V
AC
)!
c) Turn the machine on but, when “Select Program” appears,
DO NOT
push “Dialysis’ yet!
d) At the distribution board, have
Person #1
(you) HOLD the meter leads between the Heater
Connector’s BROWN and BLUE wires.
e)
Person #2:
Step #1
Select ‘Dialysis’;
Step #2
Saying “Now” press ‘CONFIRM’ to start Dialysis
Program.
f) For the
FIRST EIGHT (8) SECONDS
, after hearing “Now”, more than 100.0 volts AC measured?
Yes
More than 100.0 volts! It appears the Power Control board (inside the power supply) is bad.
No Less than 10.0 volts! TWO (2) possible scenarios below:
1)
IF (and ONLY if) “HEATER RELAY TEST FAIL” occurred:
See (ABOVE) procedure
page 578
).
2)
IF “HEATER RELAY TEST FAIL” did NOT reoccur:
See procedure number
page 583
LEFT BLANK INTENTIONALLY