
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 275
T- 2.2.1 SENSOR BOARD PIN C1 MORE THAN 9.0 VOLTS
a)
Turn the machine OFF!
b)
Set the meter to AC voltage (~ V, V
AC
)!
c) Swap in a
known good
Sensor Board (see
d) Enter Service Mode
→ Options → Hardware Options. Place the ‘X’ into the
“Yes” box and press ‘CONFIRM’. The ‘X’ turns blue!
d) If troubleshooting a DIALYSIS PROGRAM temperature problem return to it. If a HEAT DISINFECT
problem return to HEAT DISINFECT then skip to part f.
e) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
f) Figure right, measure again between the Heater
Connector’s BROWN and BLUE wires! More than 100
volts AC now (Yes or No)?
Yes
More than 100.0 volts AC! The previous Sensor
Board is bad!
IMPORTANT!
Return NTC #2’s
connector to the
1
st
distribution board position
from the left, “CON-NTC”.
No
Less than 100.0 volts AC! The heater is still off! See parts a THROUGH c below:
a)
Set the volt meter to measure DC voltage (V
DC)
.
b)
Connect the meter’s black lead to chassis ground.
c) Measure again at the Sensor Board’s motherboard connector pin C1 (VERY front pin).
1)
IF (and ONLY if) 3.6 volts DC or less:
This is a good signal so the previous
Sensor Board is bad but the heater off problem still exists. Proceed to
page 277
2)
IF more than 3.6 volts DC:
This signal continues to be ‘bad’ so the previous
Sensor Board may be good!
See parts a THROUGH d below:
a)
Turn the machine OFF!
b)
Set the volt meter to measure RESISTANCE (Ω).
Parts c and d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...