2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 588
DIASAFE- 5.0.0 ATMP STAYS 300 OR MORE / ISOLATE DIASAFE FILTER #92
a) Figure below, remove FILTER #92 from Valve #28’s tubing.
DiaSafe
®
Filter / Valve #28 / Filter #92
b) Press ‘Esc’ to return to the Home screen.
c) ENSURE no air flowing through the external flow indicator.
d)
Reset ALL alarms.
e) Select the
[Test & Options]
tab.
f) Select the ‘Diasafe Test’ button then ‘CONFIRM’ to start the test.
g) Allow the test to finish! TWO (2) possible scenarios:
1)
IF (and ONLY if) “Diasafe Test Passed”:
FILTER #92
MAY be restricted! Repeat the Diasafe
®
test once more as the failure may be intermittent!
2) IF “Diasafe Test Failed”: