2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 140
BLOCK LEAK- 2.0.0 HYDROCHAMBER LEAK LOCATED
a) Return the float’s connector to distribution board position “FLOAT-SW”.
b)
IMPORTANT!
Unclamp the Vent Tubing.
c)
Per the Figure below
, based on where a leak was seen, THREE (3) possible scenarios 1) or 2) or
3) below:
1)
IF (and ONLY if) a tubing leak:
Repair (tie wrap, if possible) and retest for a leak.
2) IF (and ONLY if) from Valve #39:
THREE (3) possibilities
1)
Bad Valve #39 mounting bracket;
2)
Bad O-ring under Valve #39;
3)
Cracked Valve #39 body.
3)
ALL OTHER locations:
Per the Figure below, if the plug is leaking replace its O-ring. All other
leaks, it may be necessary to replace the Hydrochamber.