2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 60
1)
IF (and ONLY if) the plunger moves with NO resistance:
Valve #30 is open!
See procedure
(page 60).
2) IF the plunger OFFERS RESISTANCE:
Valve #30 is restricted!
TWO (2) possible bad
components:
1)
Bad Valve #30 OR;
2)
Bad Actuator-Test Board.
To LOCATE the board refer to
(page 9).
F- 6.4.0 VALVE #30 IS OPEN
a) Reconnect valve #30’s INPUT tubing but
DO NOT
reattach the drain tubing to the rear the machine!
b) Obtain a 1000 ml graduated cylinder to measure flow subsequently.
c) Place a bucket at the drain nozzle to capture subsequent potential flow.
d) Turn Dialysate Flow ON (Flow on/off lamp NOT blinking).
e) From the Home screen, ENSURE [Dialysate Flow] is set to 800 ml/min!
f) DO NOT turn flow off again to ENSURE measurement accuracy below!
g) TWO (2) possible TMP window scenarios:
1)
IF (and ONLY if) the TMP window is NOT red:
(page 60)
2)
IF the TMP window is red (TMP alarm):
See parts A THROUGH C below:
A) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
B) Allow thirty (30) seconds.
C) If a TMP alarm reoccurs attempt RESET up to twice more BEFORE continuing to procedure
(page 60).
F- 6.4.2 ISOLATE VALVE #30 OUTPUT FLOW
a) ALLOW one (1) minute BEFORE continuing to part b
!
b) Measure flow from the REAR ‘To Drain’ port for one (1) minute. TWO (2) possible scenarios:
1)
IF (and ONLY if) LESS THAN 725 ml:
Proceed to
page 62
, procedure number
2) IF MORE THAN 725 ml:
Check the drain tubing and it’s (if used) the‘Quick Connector’ for possible
bio-growth restrictions.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...