2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 529
Yes
Strong pulses! See procedure number
(page 529).
No
Very weak or no pulses! See parts a THROUGH c below:
a) Turn the machine OFF.
b) Turn the machine on and return to Dialysis Program (
“Select Program” → ‘Dialysis’ →
‘
CONFIRM’)!
c) Proceed to
page 503
AUTOTEST- 3.0.0 DIALYSATE PRESSURE TEST
This is NOT a routine calibration. Follow the instructions exactly to avoid error!
a) Select the ‘Options’ key
→ Calibrate Sensors →
Dialysate Pressure. The screen says
“1.
Connect a pressure gauge in line…
”
b) Figure right, connect the Four-Way Assembly
into the dialyzer lines BUT
DO NOT
close
the shunt door yet!
c) A transducer protector
MUST
NOT
be installed in
the ‘to syringe’ tubing segment!
d) Place the Four-Way Assembly at dialyzer level.
e)
IMPORTANT!
Clamp the ‘to meter’ tubing
segment!
f) If using a
NEO-2
attach to the +Port (top (red) port). If using a
90XL
attach to the Pressure Module’s
Gauge Port.
g) DO NOT allow tension in the Four-Way’s tubing segments!
h)
IMPORTANT!
CLOSE the shunt door AND ENSURE
the external flow indicator’s ‘bob’ is moving at
least ¼ way up in the sight tube.
i)
Press ‘CONFIRM’. The screen says
“3. Press the [Dialysate Flow on/off] key”.
j)
Press the Dialysate Flow on/off key to turn flow off (Flow on/off lamp blinks).
k) ENSURE the external flow indicator ‘bob’ is NOT moving (i.e. Flow is off).
l)
IMPORTANT!
Remove ALL clamps!
m) Using the syringe, adjust pressure until the external meter reads 0 +/- 4 mmHg.
n) Press ‘CONFIRM’ then see procedure number
(page 530).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...