2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 412
FIL- 1.0.1 ISOLATE AIR REMOVAL SYSTEM
a) Turn the machine OFF!
b) The Deaeration gauge is used next.
ENSURE
it reads 0 inHg before installing it!
c) Figure below, tee the gauge into the Inlet (clear tubing) side of the Deaeration Pump!
d) Turn the machine on and return to Dialysis Program (“Select Prog
ram → “Dialysis’ → ‘CONFIRM).
e) Is Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes
Deaeration Pressure OKAY! See procedure number
(page 412).
No
Deaeration Pressure in NOT OKAY! ENSURING a “No Water” alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
page 498
FIL- 1.0.2 DEAERATION PRESSURE OKAY
a) Turn the machine OFF.
b) Remove the gauge and reattach the tubing.
c) Turn the machine on and
return to Dialysis Program (“Select Program → “Dialysis’ → ‘CONFIRM’).
(page 413).