2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 356
Yes
[Conductivity]
box pale yellow! See procedure number
(page 356).
No
[Conductivity]
box remains gray! Possible problem with the Cond Cell #7 circuit.
Proceed to
page 363
, procedure number
CO- 6.3.0 COND CELL #7 STABILIZES
a) Select the
[Conductivity]
data box, it turns bright yellow.
b) Enter the external meter’s conductivity.
c) ‘Sharply’ press ‘CONFIRM’ to turn
[Conductivity]
pale yellow / white. Does it remain /- 0.2 of
the external meter?
Yes
[Conductivity]
remains /- 0.2 of the meter! See procedure number
No
Read this step BEFORE performing it!
Acid Clean the machine THEN return to (ABOVE)
(page 354) HOWEVER, if you return here, swap the listed
components (see COMPONENT LIST below) one at a time in between returning to
(ABOVE) procedure number
to test each new component.
COMPONENT LIST:
1)
Conductivity Cell #7 (to LOCATE Cell #7 refer to the Figure
previous page);
2)
Sensor Board; 3
)
Functional Board (IC 20?).
CO- 6.3.2 ‘CONFIRM’ THE [CONDUCTIVITY[ DATA BOX
‘Sharply’ press ‘CONFIRM’. Does the screen say “
7. Connect the Lines to a Large Dialyzer…”
?
Yes
“
Connect the Lines to a Large Dialyzer”
(page
No a) If (and ONLY if) the screen says
“8. Reading first point for post conductivity cell.”
see
procedure number CO- 6.4.0 (page 357).
b) ‘Sharply press ‘CONFIRM’ again to save the calibration. Does any type of an “Error” banner
occur (Yes or No)?
Yes
“Error” banner occurs!.
Read before performing!
Acid clean the machine then
return to (ABOVE) procedure number
(page 354) HOWEVER, if you
RETURN HERE swap the listed components (see COMPONENT LIST below) one
at a time and repeat procedure number
COMPONENT LIST:
1)
Conductivity Cell #7 (see
2)
Sensor Board;
3)
Functional Board (IC 20?).
No a) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
b) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM!
c) Allow six (6) minutes before continuing to part d!
d) If you started out troubleshooting an OLC problem proceed to
page 386
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...