T- 1.3.90 ISOLATE NTC #2 / TEMP CONTROL
a) R
eturn to Calibrate Sensors → Temp Control
b) Press ‘CONFIRM’ to turn the
[TEMP DAC]
data box pale yellow / white.
c) ENSURING the shunt door is CLOSED the external flow indicator’s ‘bob’ MUST be moving up and down!
d) WITHOUT LOOKING AWAY, watch the external meter for seven (7) minutes OR until if it EVER
exceeds 39°.0 C even if only once. TWO (2) possible scenarios:
1)
IF (and ONLY if) the meter NEVER exceeds 39.0° C!
Press ‘CONFIRM’ twice to save the
calibration THEN proceed to
page 302
2) IF the meter exceeds 39.0° C!
TWO (2) possible scenarios i or ii below:
i.
IF NTC #2 was NOT replaced in THIS troubleshooting session:
See parts a and b below:
a) Replace
NTC #2
* with a known good. *To LOCATE NTC #2 refer to
218).
(page 243)
ii.
IF NTC #2 WAS replaced in THIS session:
TWO (2) possible bad components (see
COMPONENT LIST below). Replace each, with known good, one at a time and in between
repeat procedure number
(page 243) to test the new component.
COMPONENT LIST:
1)
Sensor Board
1,2
;
2)
Functional Board
1,2
1
To LOCATE these boards refer to
(page 9)
2
To prevent “Cond Offset Failure”
place the machine into
T and C Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
))
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...