2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 389
Step # 2: Read before parts a THROUGH e BEFORE performing them:
a) Turn the machine OFF then back on.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes ENSURING [Conductivity] stabilizes to between 13.0 and
14.3 mS.
d) Reset all alarms!
e) Return to (ABOVE)
page 385
, procedure number
HOWEVER, if
(and ONLY if) you return here continue to step #3.
Step #3
:
Read parts a THROUGH e below BEFORE performing them!
a) Perform an Acid Clean per procedure.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes ENSURING [Conductivity] stabilizes to between 13.0 and 14.3
mS.
d) Reset all alarms!
e) Return to (ABOVE)
page 385
, procedure number
HOWEVER, if you
return here (because
“Can’t Run OLC Test”
reoccurs) proceed to
page 354
procedure number
to calibrate COND CELLS. After the calibration if (and
ONLY if) you return here, THREE (3) possible bad components below:
Component List:
1)
Power Logic Board; 2) Actuator-Test Board;
3)
Functional Board.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...