2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 565
1)
IF (and ONLY if), during this procedure, tubing ‘blew off’ the DiaSafe
®
filter or the
top side Balancing Chamber Valves (see Figure below):
A)
Turn the machine off;
B)
Reattach and tie wrap all tubing;
c)
Repeat (ABOVE) procedure number
(page 564).
2) IF tubing DID NOT ‘blow off’:
A)
Reconnect the drain tubing;
B)
NOTE
ONLY
VALVE
#25
will be checked and proceed to
page 192
LEAKING 5.2.0 VALVE #25 OKAY / ISOLATE VALVE #26
a) Press ‘Escape’ then ‘CONFIRM’ twice to call the “Select Program” screen!
b) Remove the red dialyzer connector from the shunt door and place it on the floor!
c) Close the door!
d) Per the Figure above, clamp and remove Valve #26’s
INPUT tubing
!
e) Fill the 60 ml syringe with water then attach it to Valve #26’s INPUT nozzle.
f) Push
AS HARD AS YOU CAN
on the syringe plunger. Can you push ALL 60 ml through Valve #26 within
FIVE (5) seconds?
Yes
Valve #26 is okay! Return the dialyzer connector to the shunt then see procedure number
(page 566).
No Significant resistance felt!
NOTE
ONLY
VALVE #26
will be checked and proceed to
page 192
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...