2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 439
FIL- 7.0.0 ISOLATE AIR LEAK
a) Remove the Four-Way.
b)
IMPORTANT! Return the Dialyzer lines to the shunt door!
c) Turn the machine OFF!
d) Turn the machine on and return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
e) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
f) Allow [
Conductivity
] to increase to more than 13.0 mS!
g) Figure below, is air still seen at Chamber #69?
Yes Air seen! Proceed to
page 490
SECTION 11 - INDUCED AIR LEAK TESTS
No air seen! Call debug screen 1. WITHOUT LOOKING AWAY watch
FILACT
(middle column)
for up to two (2) minutes. If EVER = 1, even just once, indicates a Filling Program!
TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT EVER = 1:
Return to ABOVE procedure number
page 421
2) IF FILACT ALWAYS = 0:
Problem solved!