2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 482
i)
IF FILACT EVER ‘1’ (even if only once):
. Proceed to
page 409
ii)
IF FILACT ALWAYS ‘0’!
Proceed to
page 440
, SECTION 9 – TMP (PRESSURE)
PROBLEMS.
PHT- 3.1.3 PRESSURE TEST FAILS AND ATMP INCREASED TO MORE THAN -400
a) Remove the UF Sample Connector (Figure right) AND
hold it so that you can observe UF PUMP output!
b) Below you will
simultaneously
watch the screen
banners, debug screen 1’s
ATMP
AND the UF Sample
Connector!
c) Reset all alarms. From the Test & Options screen select
the [Pressure Tests] button then press ‘CONFIRM’.
d) During “Get Neg TMP” the UF Pump sends ‘strong pulses’ through the Sample Connector every eight
(8) seconds until the “Neg Flow On” banner appears THEN the UF Pump SHOULD stop!
e) ENSURING “Neg Flow On” is presenting AND
ATMP
is REMAINING more than -270 does the UF
Pump CONTINUE to send strong pulses(s) through the UF Sample Connector?
Yes If ABSOLUTELY SURE the UF Pump sent a pulse WHILE the “Neg Flow On” banner was up!
Turn the machine off then see the TWO (2) possibilities (below). Repeat the Pressure Tests
to see if either solved the problem.
Possibilities
:
1)
A card cage circuit board not seated properly OR
2)
Bad Actuator-Test
Board.
No
If the Pressure Tests Fail, from the Tests & Options screen, if a Negative Pressure Test listing
appears see procedure number
(page 482). If a Positive Pressure Test listing
appears return to Scenario #4 of ABOVE procedure number
(page 481). If (and
ONLY if) the Pressure Tests pass proceed to
page 484
, procedure number
PHT- 3.1.4 NEGATIVE PRESSURE TESTS FAIL
1) IF (and ONLY if) “Fail Neg Stabilize” is listed.
TWO (2) possible scenarios i) or ii) below:
i)
IF FILACT EVER ‘1’ (even if only once) during the repeat 30 second STABILIZATION
“Neg Flow On” period:
Proceed to
page 409
ii)
IF FILACT ALWAYS = 0 during the repeat 30 second “Neg Flow On” STABILIZATION
period:
Proceed to
page 440
SECTION 9 – TMP (PRESSURE) PROBLEMS
2) IF “Fail Neg Flow On” is listed:
TWO (2) possible scenarios i) or ii) below:
i)
IF FILACT EVER ‘1’ (even if only once) during the repeat 30 second HOLD “Neg Flow On”
period:
Proceed to
page 409
SECTION 8 – FILLING PROGRAM PROBLEMS
ii)
IF FILACT ALWAYS = 0 during the repeat 30 second HOLD “Neg Flow On” period:
Proceed to
page 440
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...