2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 516
BL- 4.0.0 NO AIR SEEN / ISOLATE BLOOD LEAK DETECTOR #8
a) If automated tests are running (screen reads “Test: …) allow them to finish before
continuing!
b) Call debug screen 4 to see
LEAK
and
DIMN
. FIVE (5) possible scenarios, 1) or 2) or
3) or 4) or 5) below:
1)
IF LEAK is between 4.5 and 5.2 AND DIMN between 4.0 and 6.0:
This is normal! Acid clean
then Bleach Disinfect. If (and ONLY if) a false Blood Leak reoccurs in the near future see
procedure number
page 520
).
2) IF LEAK AND / OR DIMN are less than 0.2:
3)
IF LEAK is more than 5.2; DIMN does not matter:
page
4) IF DIMN is more than 6.0; LEAK does not matter:
page 517
5) All OTHER scenarios:
page 520
BL- 4.1.0 ISOLATE ‘OPEN CIRCUIT’ BLOOD LEAK DETECTOR
a)
IMPORTANT! To prevent damage turn
the machine off!
b) Figure right, at the distribution board, ENSURE the
Blood Leak Cable is plugged in properly.
If not, this
may be the problem!
c) A procedure, in different Section, is performed next.
IMPORTANT!
NOTE
this page and procedure
number (
) because you may prompted to return to here.
d) BEFORE continuing to part e the Sensor Board cable MAY be bad.
NOTE
that five (5) BLOOD LEAK
connections will be checked and proceed to
page 524
, to perform
e) If the Sensor Board cable checks okay return to Dialysis Program (
“Select Program” → ‘Dialysis’ →
‘CONFIRM’)!
f)
Allow six (6) minutes for stabilization!
g) ENSURING the external flow indicator’s ‘bob’ is rising and falling see procedure number
(
page 518
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...