P- H.1.4 ISOLATE FUNCTIONAL BOARD DEVICES
a)
IMPORTANT! To prevent damage turn the machine OFF!
b) Device cables along the TOP EDGE of the Functional Board AND Power Logic Board will be unplugged
next.
Per the Figure below
, to prevent damage when eventually returning them,
RECORD
where
they are plugged in BEFORE continuing to part c.
c) Unplug the device cables but
DO NOT
unplug the Power Logic board’s front cable (closest to the
screen)!
NOTE:
Ignore “Blood Pump No Communication” if one occurs!
d) Repeat the Voltage Detection Calibration. THREE (3) possible scenarios 1) or 2) or 3):
1)
IF “Operator Error” reoccurs:
2)
IF 5V EST AND/OR 12V EST ar
e
not in range:
3) IF “Operator Error” did NOT occur AND 5V EST AND 12V EST are in range:
One of the devices
may be causing the problem. To locate which one perform parts A through C below:
A)
Turn the machine OFF!
B)
Your choice, return ONE of the device cables to WHERE IT BELONGS.
C)
Attempt the Voltage Detection Calibration. When the problem reoccurs the last device
plugged in or its cable is causing the problem!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...