2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 153
NW- 5.0.0 CHECK FLOAT AND VALVE #41
a)
IMPORTANT!
TURN THE WATER OFF!
b) Obtain a 1000 ml graduated cylinder.
c) Per the Figure below, locate the clear tubing (from Valve #41) that enters the ‘Air Gap Tower’ vertically
at the top of the Hydrochamber.
Figure 31 – Inlet Water Tubing
d) Pull the tubing off the ‘Air Gap Tower’ and direct it into the cylinder.
e) TURN THE WATER ON and measure for one (1) minute!
f)
TWO (2) possible scenarios:
1)
IF (and ONLY if) more than 800 ml collected:
(page 154).
2)
IF less than 800 ml collected:
There is not enough water from the RO OR a there is a restriction
at, to or from Valve #41.
.