2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 144
NW- 1.0.5 “NO WATER” IN HEAT DISINFECT / ISOLATE RINSE PROGRAM
a) Press ‘Escape’ TWICE then ‘CONFIRM’ TWICE to call the “Select Program” screen!
b)
Place the machine into RINSE (NOT HEAT DISINFECT)!
c) Allow one (1) minute BEFORE continuing!
d) WITHOUT LOOKING AWAY, watch for four (4) minutes OR until if a “No Water” alarm EVER reoccurs?
Yes “No Water” reoccurs! Proceed to
page 146
No “No Water” does NOT reoccur! The alarm may be related to Heat Disinfect ONLY! See
(page 144).
NW- 1.0.6 IN RINSE “NO WATER” DOES NOT REOCCUR
a) Press ‘Escape’ then ‘CONFIRM’ twice to call the “Select Program” screen.
b)
Place the machine in HEAT DISINFECT (NOT RINSE)!
c)
Open the shunt door and LEAVE IT OPEN until instructed!
d) Figure right, from the Heat Disinfect screen,
does the TOP
Remaining Prerinse Time
window = 0:00 min:sec (Yes or No)?
Yes
Remaining Prerinse Time
= 0:00! A
“No Water” alarm should not be
occurring unless the water is off!
No
Remaining Prerinse Time
does NOT
NW- 1.0.7 REMAINING PRERINSE DOES NOT = 0:00
WITHOUT LOOKING AWAY, watch for four (4) FULL minutes OR until if a “No Water” alarm EVER
reoccurs?
Yes
“No Water” alarm reoccurs! See procedure number
No
“No Water” DOES NOT reoccur! DO NOT continue but close the shunt door!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...