2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 178
Yes
More than 300 ml! Valve #24 is open! Without returning the connector to the shunt, see
procedure number
(page 179).
No
Less than 300 ml! ENSURING the “Interrupted” banner IS NOT up, see parts a THROUGH c
below:
a) Return the dialyzer connector to the shunt but
DO NOT
close the door!
b)
Per the Figure below
, routing Valve #24’s
INPUT
tubing so that spillage into the
hydraulics does NOT occur, REMOVE the tubing and measure from it for thirty (30)
seconds.
c) More than three-hundred (300) ml every thirty (30) seconds?
Yes
More than 300 ml!
A)
Return Valve #24’s tubing.
B)
NOTE
ONLY
VALVE
#24 will be checked and proceed to
page 192
No Less than 300 ml! See parts a AND b below:
a) Return Valve #24’s tubing.
b) If Loading AND Flow Pressure are still good, FOUR (4) possible bad
components:
1)
Restricted DiaSafe
®
filter;
2)
Bad Actuator-Test Board OR
3)
Bad Actuator Cable; OR
4)
Multiple stuck closed balancing chamber
valves.