2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 459
TMP- 2.2.8 -250 mmHg HOLDING / ISOLATE DIALYSATE PRESSURE (PDIAL)
a) ENSURING the external meter remains about -250 mmHg, call
debug screen 0.
b) Figure right, if (and ONLY if) Valve #25’s ‘dot’ is BLUE continue to part
c. If NOT blue, return to (ABOVE) procedure number
(
page 456
c) Figure right, look at screen 0’s
PDial
data
window. TWO (2)
possible scenarios:
1)
IF (and ONLY if) PDial AGREES with the external meter /- 35:
See procedure number
(page 459).
2)
IF PDial DOES NOT agree with the external meter /- 35:
Proceed to
page 470
TMP- 2.3.0 CHECK PDIAL TO 520 mmHg
a) PULL on the syringe plunger to create negative pressure.
b) Clamp the ‘to syringe’ tubing segment.
c) Remove the syringe then reattach it.
d) PULL again. Can you make screen 0’s
PDial
reach
at least
-
520 mmHg?
Yes
-520 at
PDial!
(page 459).
No Cannot achieve -520! Proceed to
page 470
, procedure number
TMP- 2.3.1 CHECK PDIAL ZERO
a) Remove the syringe and all clamps to open the four way to atmosphere!
b) ENSURING the transducer protector, at the meter, is NOT wet, does screen 0’s
PDial =
0 +/- 40?
Yes
PDial
= 0 +/- 40! See procedure number
No
PDial
is NOT
between 0 and 40! Proceed to
page 470
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...