2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 172
CLEAN- 3.0.0 LOADING PRESSURE OKAY / ISOLATE MOTORS
a)
If the “Interrupted” banner is on the screen return to Heat Disinfect OR Rinse!
b) THREE (3) checks on the Flow AND Deaeration motors (Figure below):
Check #1:
Is the FLOW MOTOR shaft rotating in the same direction as the Deaeration Motor i.e.
counterclockwise (CCW)?
Yes
Flow Motor rotating CCW! See CHECK #2.
No If (and ONLY if) it is NOT rotating CCW,
NOTING
ONLY the FLOW MOTOR
will be checked proceed to
page 130
Check #2:
Using the handle end of a screwdriver, push hard on the FLOW MOTOR’S shaft, release it,
and then push again several times. Can you make stop rotating and REMAIN stopped?
Yes
If ABSOLUTELY SURE the motor stops! TWO (2) possible bad
components:
1)
Bad Flow Motor (probably brushes) OR;
2)
Bad Flow
Pump head.
No Flow Motor does NOT stop! See CHECK #3.
Check #3:
Push hard on and release the DEAERATION MOTOR shaft. Can you make it stop rotating?
Yes
If ABSOLUTELY SURE the motor stops! TWO possible bad components:
1)
Bad Deaeration Motor (possibly brushes) OR;
2)
Bad Deaeration Pump
head.
No
The motor does NOT stop! Perform parts a AND b below:
a) Based upon if a debug screen is up, TWO (2) possible scenarios:
1)
IF a debug screen is up:
Call “Select Program” by pressing ‘Escape’
TWICE then ‘CONFIRM’ TWICE.
2)
IF a debug screen is not up:
Call “Select Program” by pressing
‘Escape’ ONCE then ‘CONFIRM’ TWICE.
b)
(page 173).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...