2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 656
DVF- 2.1.0 ISOLATE VALVE #24 SHORT CIRCUIT
a) ENSURE a “No Water” alarm NEVER appears!
b) Call debug screen 0. WITHOUT LOOKING AWAY, watch
Flow Error
,
for three (3) minutes. If EVER = 1, even if only once, indicates a Flow
Error. TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error EVER = 1:
Return to
page 27
, procedure number
2) IF Flow Error ALWAYS = 0: P
erform parts A THROUGH I below
A) Call the Home screen.
B) Allow
Temperature
to reach between 35.0 and 39.0 °C;
Conductivity
between 13.0 and 14.5
mS!
C) At the bottom of the screen, select the ‘Dialysate’ tab.
D) If necessary, adjust the Conductivity Limits until ‘Actual’
Conductivity is CENTERED between the Limits.
E) Press ‘CONFIRM’
F) Call debug screen 0 to locate Valve #24’s ‘dot’ (Figure right).
G)
CLOSE THE SHUNT DOOR!
H) Allow up to two (2) minutes OR until Valve #24’s ‘dot’ turns blue.
I) Allow five (5) MORE minutes! Does “Dialyze Valve Fail” OR “Act Byp Valve Fail” alarm banner
occur?
Yes
“Dialyze Valve Fail” OR a “Act Byp Valve Fail” occurs! Electrical problem with
Valve #24. Proceed to
page 657
, procedure number
No Whatever was causing the problem is no longer present! Watch for ten (10 minutes).
If a Flow Error reoccurs proceed to
page 27
, procedure number
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...