2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 627
P- H.4.8.2 REMAINS STABLE AND NO ALARMS OCCUR / HEAT-RELATED PROBLEM?
a) Place the machine into Heat Disinfect then call debug screen 10 to watch the
24V
window.
b) WITHOUT LOOKING AWAY, watch the volt meter until if the program completes.
c) Does the measurement become unstable AND/OR does a “24V” OR “Failed Sending Data To The
Actuator Board” alarm occur before Heat Disinfect completes?
Yes
Unstable measurement OR “24V” OR “Failed Sending Data…” alarm! See procedure
number
(page 627).
No Return to Dialysis Program. Does a “24V” OR “Failed Sending Data To The Actuator Board”
alarm occur (Yes or No)?
Yes
One at a time, starting with the Actuator-Test Board, swap in the card cage
circuit boards, with known good then, in between, repeat procedure number
(page 627) to test each new board.
No The Troubleshooting Guide cannot locate an immediate problem!
P- H.4.9 UNSTABLE OR ALARM REOCCURS / ANALYZE PROBLEM
TWO (2) possible scenarios, based on MEASURED voltage:
1)
IF (and ONLY if) MEASURED voltage became UNSTABLE:
Proceed to
page 649
, procedure
2) IF MEASURED voltage remained STABLE:
Perform parts a THROUGH c below:
a)
Place the machine into Service Mode → Options → Hardware Options.
b) Next to
T and C Mode
place the ‘X’ in the “Yes” box and press ‘CONFIRM.’
The ‘X’ turns blue.
c) With the machine off, swap in the card cage circuit boards, with known good, one at a time, starting
with the Actuator-Test Board and, in between, return to Heat Disinfect to test each new board. If
“24V” OR “Failed Sending Data…” alarm reoccurs swap in the next board until these alarms DO
NOT occur.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...