2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 340
CO- 2.0.8 [CONDUCTIVITY] WINDOW IS WHITE / ISOLATE ‘OUT OF BYPASS’ FLOW
A)
IMPORTANT!
If not already, close the shunt door!
B) From the Home screen, ENSURE [Dialysate Flow] remains set to at least 500 ml/min!
C)
DO NOT reset alarms!
D) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! Proceed to
page 342
, procedure number
No
See part E.
E) Call debug screen 2. To ENSURE the shunt door is really closed
CVRCLS
(2
nd
column from left) = 1!
F) Call debug screen 0. Allowing up to three (3) additional
minutes, does Valve #24’s ‘dot’ (Figure right) turn BLUE?
Yes
Valve #24’s dot’ is BLUE! See procedure
number
(page 341).
No
After 3 minutes Valve #24’s ‘dot’ stays WHITE!
See parts a THROUGH d below:
a) If removed from a previous
procedure, return Conductivity
Cell #7’s connector to distribution
board position “X7, COND”
(Figure right).
b) Return the concentrate
connectors to their rinse ports.
c)
IMPORTANT! Place
the machine into
RINSE!
d) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! Proceed to
page 353
, procedure number
No ‘Bob’ NOT moving! ENSURING the machine is in RINSE AND the bob is
NOT moving, see parts a AND b below:
a) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ →
‘CONFIRM’)!
b) Proceed to
page 20
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...