2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 219
T- 1.0.6 OPTIONAL BLOOD TEMPERATURE MODULE (BTM) EQUIPPED?
Per the Figure right, TWO (2) possible scenarios:
1)
IF NOT equipped with the optional
Blood Temperature Module (BTM):
See
2)
IF equipped with a Blood Temperature
Module (BTM):
See parts a THROUGH g
below:
a) Place the machine into Serv
ice Mode → Options → Hardware Options.
b) Next to ‘
BTM
’, place the ‘X’ into the “No” box.
c) Press ‘CONFIRM’ to turn the “X” blue!
d)
To prevent damage turn the machine OFF!
d) To avoid pulling cables loose, GENTLY open the card cage.
e)
f) The Functional Board is the 3
rd
circuit board from the right! Figure below, unplug the ten (10) pin
ribbon cable from the P8 connector.
g)
Close the card cage!
h) Where was the problem ORIGINALLY occurring,
1) In Dialysis Program i.e. connected to
concentrate
? OR
2) In Heat Disinfect
? TWO (2) possible scenarios:
1)
IF (and ONLY if) in Dialysis Program
:
(page 221).
2) IF in Heat Disinfect
:
Return to Heat Disinfect and allow up to thirty (30) minutes. TWO (2)
possible scenarios next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...