2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 511
Do NOT troubleshoot Blood Leak if a “No Water”, Flow Error, Temperature, or Conductivity
alarm EVER occur! The machine MUST NOT be ‘in bypass’!
SECTION 15 - BLOOD LEAK PROBLEMS
The Blood Leak Detector #8 detects blood as little as 0.35 ml / Liter! If a patient is on the machine, before
considering service, check for a ‘real’ blood leak using a Hemastix
®
strip or equivalent per clinic procedure.
A)
ENSURE the dialyzer connectors are connected PROPERLY to the shunt door!
B)
CLOSE THE SHUNT DOOR!
C)
DO NOT
reset alarms until instructed!
D)
[
Temperature
] MUST be between 35.0 and 39.0 °C; [
Conductivity
] between 13.2 and 14.5 mS. BOTH
MUST BE STABLE i.e. NOT changing more than +/- 0.2 per minute!
E)
If automated tests EVER start (screen reads “Test:.…) allow them to finish BEFORE continuing!
F)
From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
G) Call debug screen 0 to see
Flow Error
. If it EVER = 1, even once, indicates a masked Flow Error!
WITHOUT LOOKING AWAY, watch
Flow Error
for
two (2) minutes or until if it EVER = 1! TWO (2)
possible scenarios 1) or 2) below:
Yes
Flow Error EVER = 1, even if only once:
Proceed to
page 20
No
Flow Error ALWAYS = 0:
See parts a AND b below:
a) Valve #24’s ‘dot’ MUST be BLUE i.e. machine is not in bypass BEFORE continuing to part b!
b) Call the Home screen. Figure right, what color is the TMP Window?
1)
IF (and ONLY if) pale yellow/white:
See procedure number
(page 511).
2)
IF RED (TMP alarm present):
A)
Press and release the ‘RESET’ key then immediately
press and hold it for three (3) seconds;
B)
Allow thirty (30) seconds;
C)
If a TMP alarm
reoccurs, attempt RESET up to twice more BEFORE continuing to procedure number
(page 511).
BL- 1.0.4 ISOLATE TMP
TMP
is STABLE ONLY if the
TMP
Window stays white AND its value does NOT change more than +/- 60
mmHg in three (3) minutes! TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is UNSTABLE:
Proceed to
page
2) IF TMP is STABLE:
.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...