
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 625
P- H.4.5 +24 VOLT- A TROUBLESHOOTING
a) Spread the card cage panels open then gently drop the front panel down.
b)
ENSURE the volt meter’s black lead remains attached to chassis ground!
c) ENSURE the machine in on (fan running)!
d)
below
, place the red meter lead into the RIGHT SIDE socket of the motherboard’s white, two
(2) pin, +24V-A connector. BE CERTAIN to make good contact with the metal pin inside the socket!
e) TWO (2) possible scenarios:
1)
IF between 23.0 and 28.0 volts DC (V
DC
):
(page 626).
2)
IF NOT between 23.0 and 28.0 volts DC (V
DC
):
Proceed to
page 649
, procedure number
Figure 119 – Mothe24V-A Connector
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...