2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 566
LEAKING- 5.3.0 VALVE #26 OKAY / ISOLATE VALVE #29
a)
IMPORTANT!
Reattach Valve #26’ its tubing!
b) Figure right, screw the 60 ml syringe, filled with water,
onto the Fluid Sample Port.
c) This procedure is time sensitive! Read parts d through
g BEFORE performing them!!
d)
IMPORTANT!
Return to RINSE Program!
e) Call debug screen 0. If the debug screens do not
appear press ‘Escape’ then call screen 0.
f) Figure right, locate Valve #29’s ‘dot’. It cycles from
white to blue every two (2) minutes and stays blue
(valve open) for nine (9) seconds.
g) When Valve #29’s ‘dot’
FIRST
turns BLUE push
HARD
on the plunger! You should feel very little resistance!
h) Were you able to push water through the Sample Port
WHILE Valve #29’s ‘dot’ is blue?
Yes
All water pushed out! Flow through Valve #29!
See procedure number
(page 567).
No
Significant resistance felt!
NOTE
ONLY
VALVE #29
(Figure below) will be checked and
proceed to
page 192