2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 567
LEAKING- 5.5.0 VALVE #29 IS OKAY / ISOLATE FOR A ‘PARTIAL’RESTRICTION
a)
IMPORTANT! Turn the machine OFF then back on.
b) If the screen’s Heat Disinfect button is BLUE go to Heat Disinfect. If gray go to
RINSE
!
c) Allow ONE (1) FULL minute for debug screens to update!
d) Call debug screen 10 (Figure right)
to locate
PDIA
e) WITHOUT LOOKING AWAY, watch
PDIA
for FIVE (5) full minutes. It is
NORMAL it cycle between 0.0 (for
about three seconds to more than
0.1 for about one (1) second. It
should NEVER be more than 6.0!
THREE (3) possible scenarios 1) or
2) or 3) below:
1)
IF (and ONLY if) PDIA CYCLES from between 0.0 to more than 0.1
BUT NEVER more than
6.0l
! A problem is not present at this time. Tie wrap or replace ALL tubing connections that were
leaking and ENSURE the machine passes the alarms and pressure tests!
2) IF (and ONLY if) PDIA REMAINS less than 1.0 for longer than 10 seconds:
If a Flow Error
occurs OR a leak reoccurs there MAY be a partial restriction to the drain (Valve #30, Valve #29).
3)
IF PDIA goes to more than 6.0 AND a Flow Error occurs:
Referring to the Figures (below)
Valve #24 OR #25 is partially restricted!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...