2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 112
F- 14.0.0 ISOLATE 'OUT OF BYPASS' CIRCUIT
a) Turn the machine OFF!
b) Figure right, ENSURE the flow indicator’s INNER Tapered
Tube AND its ‘bob’ is orientated narrow to wide from the
bottom up.
(page 112).
F- 14.0.2 ISOLATE VALVE #24 / TO DIALYZER CIRCUIT
a) Obtain a 1000 ml graduated cylinder!
b)
Service Mode’s Flow Pressure is used to isolate the ‘out of bypass’ circuit. DO
NOT follow the screen instructions! Instead following the procedures below!
c) E
nter Service Mode → Calibrate Hydraulics → Flow Pressure.
d) Press ‘CONFIRM’ THEN ENSURE the FLOW MOTOR shaft is rotating!
e) Remove the BLUE dialyzer connector from the shunt door and measure from it for one (1) minute.
f) Press ‘CONFIRM’ to return to “Calibrate Hydraulics”! TWO (2) possible scenarios:
1)
IF (and ONLY if) MORE THAN 800 ml collected!
Valve #24 is okay! See procedure number
(page 112).
2)
IF LESS than 800 ml collected!
Bad flow from Valve #24. Proceed to
page 114
, procedure
F- 14.0.4 VALVE #24 IS OKAY / ISOLATE VALVE #25 / FROM DIALYZER CIRCUIT
a) Return the BLUE connector to the shunt and close the door.
b) Return to
Calibrate Hydraulics → Flow Pressure.
c) Press ‘CONFIRM’ THEN ENSURE the FLOW MOTOR
is rotating!
d) Figure right, remove the Flow Pump’s clear INPUT
tubing and measure from it for one (1) minute.
e) Press ‘CONFIRM’ to return to “Calibrate Hydraulics”!
f) Reattach the tubing. TWO (2) possible scenarios:
1)
IF (and ONLY if) MORE THAN 800 ml collected!
The ‘out of bypass’ circuit checks good! See
(page 113).
2)
IF LESS THAN 800 ml collected!
A restriction in the ‘from (red) Dialyzer’ line is indicated! See
parts a THROUGH c next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...