2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 160
CLEAN- 1.2.3 ISOLATE LOADING PRESSURE (3)
A)
Loading Pressure may or may not be cycling. TWO (2) possible scenarios below:
1)
IF NOT cycling:
Pressure is ‘OKAY’ if it REMAINS between 22 and 27 psi. See part B!
2)
IF cycling:
It may cycle to but SHOULD NOT remain less than 15 psi. It is ‘OKAY’ if it EVER
cycles to a PEAK of somewhere between 22 and 27 psi. See part B!
B)
FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1)
IF (and ONLY if) Loading Pressure is ‘OKAY’:
Leaving the gauge installed, proceed
page
, procedure number
2)
IF (and ONLY if) REMAINS ALWAYS less than 15 psi:
ENSURING
the gauge was SLAMMED
into the Rinse port, Loading
Pressure is low!
(page 161).
3)
IF (and ONLY if) EVER MORE than 28 psi:
Turn Valve #65’s nut (Figure below)
counterclockwise (outward) to adjust to a PEAK of between 22 and 25 psi. If a Flow Error
reoccurs proceed
page 172
, procedure number
4)
IF cycling to a PEAK of between 15 and 21 psi:
ENSURING
the gauge was SLAMMED into
the Rinse port, turn Valve #65’s nut (Figure below) clockwise (inward) attempting to adjust to a
PEAK of between 22 and 25 psi. Can you?
Yes
Pressure CAN be adjusted to its PEAK! Allow two (2) minutes then if a Flow Error
reoccurs proceed
page 172
, procedure number
No Loading Pressure remains low! See procedure number
(page 161).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...