2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 357
CO- 6.4.0 OLC (COND CELL #13) CALIBRATION
This four-point calibration sets slope for Post Dialyzer Cond Cell #13. See
parts a THROUGH e below:
a) Figure right, connect the dialyzer lines to a PRIMED dialyzer
1,2,
.
IMPORTANT!
The dialyzer MUST positioned vertically with the
red dialyzer quick connector AT THE BOTTOM!
1
If a dialyzer is not used OLC problems may occur!
2
ENSURE the blood ports are tightly sealed to prevent air leaks!
WARNING!
Do NOT
use the blood port caps from the dialyzer’s
packaging!
Blood tubing (Figure right) is BEST!
b)
IMPORTANT! CLOSE the shunt door!
c) Allow one (1) FULL minute!
d) Press ‘CONFIRM’. The screen says “
8. Reading first point for post conductivity cell.”
e) After no more than ten (10) minutes, does the
[Conductivity]
data box stabilize between 15.00 and
15.90 mS causing the screen to advance to:
“9. Reading second point for post conductivity cell.”
?
Yes
Screen advances to the second point! See procedure number
No Screen DOES NOT advance to the second point! TWO (2) possible scenarios
below:
1)
IF (and ONLY if) “COND LESS THAN 10.0” OR “COND LESS THAN 15.0” banner
occurred:
2)
All OTHER scenarios:
See parts a AND b below:
a) If a new dialyzer was used and it was not PRIMED it may cause an unannounced
Flow Error which makes the external indicator’s ‘bob’ stop rising and falling. In this
event ‘Escape’ the calibration then reenter it.
b) If the external flow indicator’s bob is moving up and down proceed to
page 360
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...