2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 583
HEATER RELAY- 3.0.0 “HEATER RELAY TEST FAIL” INTERMITTENT
A)
Turn the machine off for two (2) seconds then back on.
B)
Return to Dialysis Program (“Select
Program” → ‘Dialysis’ → ‘CONFIRM’). If (and ONLY if) “HEATER
RELAY TEST FAIL” reoccurs see procedure number
(page 583). If “HEATER
RELAY TEST FAIL” DOES NOT occur see part C.
C)
Repeat parts A and B four (4) times. If (and ONLY if) “HEATER RELAY TEST FAIL” does not reoccur
whatever was causing the problem has cleared!
HEATER RELAY- 3.1.0
D)
Turn the machine off and swap in a known good Power Logic board. To LOCATE the board refer to
(page 9).
E)
Turn the machine on and return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’). If
(and ONLY if) “HEATER RELAY TEST FAIL” reoccurs see procedure number
(page 583) otherwise see part F.
F)
Turn the machine off then repeat parts E and F four (4) times. If “HEATER RELAY TEST FAIL” does
not reoccur the new Power Logic board fixed the problem!
HEATER RELAY- 3.2.0
NOTE:
The first Power Logic Board is probably okay.
G) Turn the machine off. THREE (3) possible bad component (see COMPONENT LIST below). Swap in
one (your choice), one at a time, then see part H to test the new component.
COMPONENT LIST: 1)
Power Control board (inside the power supply) ;
2)
Functional Board (to prevent
a “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
));
3)
Motherboard.
H) Turn the machine on and return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’). If
(and ONLY if) “HEATER RELAY TEST FAIL” reoccurs repeat part G until HEATER RELAY TEST FAIL”
does not reoccur indicating the last component swapped in fixed the problem
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...