2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 393
CO- 10.0.705 ISOLATE OLC PROBLEM
a)
IMPORTANT!
TURN the machine OFF!
b) Place the concentrate connectors into their rinse ports
c) Turn the machine on and place it into Acid Clean.
d) After rinse is complete, connect to fresh concentrate and return to Dialysis Program.
e) Allow six (6) minutes for Temperature and Conductivity to stabilize.
f)
Read before performing!
Return to (ABOVE) procedure number
(page 385)
HOWEVER, if (and ONLY if) you return here see procedure number
CO- 10.0.706 ISOLATE OLC PROBLEM
a)
Read before performing!
Proceed to (ABOVE) procedure number
(page 354) to calibrate
COND CELLS HOWEVER, if (and ONLY if) you return here see part b.
b)
Read before performing!
Swap the following components in (see COMPONENT LIST below) one at a
time, with known good then return to (ABOVE) procedure number
(page 385) to test the
new component. If (and ONLY if) you return here continue through the list until the OLC problem is
solved.
COMPONENT LIST:
1)
Power Logic Board OR; 2) Actuator-Test Board OR; 3) Functional Board
1
1
Cond Cells MUST be calibrated after the Functional board is swapped in before its OLC
functions can be tested!
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...