2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 387
CO- 10.0.44 PRE DEV VALUE LESS THAN 135 / CONDUCTIVITY STABILITY
a)
Post Dev val
(right column) is the conductivity difference between now and one (1)
minute ago from Post-Conductivity Cell #13. If no alarms it changes continually!
b) After no more than three (3) minutes, if not already,
Post Dev val
should fall to less than 70. TWO (2)
possible scenarios:
1)
IF Post Dev val is OR falls to less than 70:
(page 388).
2) IF Post Dev val NEVER falls below 70:
Post Cond is unstable!
ENSURING no alarms occurred
see parts a and b below:
a) Per the Figure above, using a flashlight and WITHOUT LOOKING AWAY, for one minute, if air
bubbles are flowing through the Post Dialyzer Cond Cell #13’s locate and repair the air leak!
b) If (and ONLY if) absolutely sure of no air proceed to
page 391
, procedure number