2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 626
P- H.4.6 BETWEEN 23.0 AND 28.0 VOLTS / MONITOR 24 VOLTS
a) With the “Select Program” banner up select the screen‘s ‘Dialysis’ button but
DO NOT
press CONFIRM
yet!
b) Call debug screen 10. Figure right,
24V
is the ‘real time’ reading from
the +24 volt supply.
HI
and
LO
‘records’ the highest and lowest
starting one (1) minute after the machine is turned on.
c) At the bottom of the screen, select the ‘Dialysate’ tab THEN press ‘CONFIRM’ to return to Dialysis
Program!
d) Call debug screen 10 and allow one (1) minute for the
24V
window to update. The
24V
window IS
NORMALLY UNSTABLE however, an alarm does not occur unless if it falls below 22.8 volts (“24V Low”)
or rises above 28.0 volts (“24V High”) for several seconds.
e) WITHOUT LOOKING AWAY, watch the volt meter for ten (10) minutes. Does the measurement become
unstable (i.e. changes more than +/- 0.5) OR does a “24V” OR “Failed Sending Data To Actuator Board”
alarm occur?
Yes
Measurement became unstable AND / OR “24V” OR “Failed Sending Data…” reoccurs!
See procedure number
(page 626).
No Measurement remains stable AND a “24V” OR “Failed Sending Data To Actuator Board”
alarm reoccurs! Proceed to
page 627
, procedure number
.
P- H.4.7.1 UNSTABLE OR 24V OR FAILED SENDING REOCCURS
Based on the MEASURED voltage, TWO (2) possible scenarios:
1)
IF MEASURED voltage became UNSTABLE:
Proceed to
page 649
, procedure number
2) IF MEASURED voltage remained STABLE:
Perform parts a through c below:
a) Place the machine into Service
Mode → Options → Hardware Options.
b) Next to
T and C Mode
place the ‘X’ in the “Yes” box and press the
‘CONFIRM’ key. The ‘X’ MUST turn blue!
c) One at a time, starting with the Actuator-Test Board, swap in the card cage circuit boards, with
known good then, in between, return to (ABOVE) procedure number
page 626)
to test
each new board.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...