![Fresenius Medical Care 2008K2 Скачать руководство пользователя страница 217](http://html.mh-extra.com/html/fresenius-medical-care/2008k2/2008k2_troubleshooting-manual_2331635217.webp)
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 213
VEC- 2.2.0 VALVE #39 IS OPEN
A) Turn the machine OFF and remove the gauge!
B) Turn the machine on and allow the “Select Program” screen to appear. What color
is the ‘Dialysis’ button (Gray or Blue)?
1) IF GRAY:
Place the machine in RINSE. Most Valve Errors will not stop Program timing for very
long. Allow RINSE to complete i.e. [Remaining Time] = 0:00, then see part C.
2) IF BLUE:
See part C.
C) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
D) From the Home screen, set [Dialysate Flow] to 800 ml/min
E) Allow five (5) minutes BEFORE continuing to part F.
F) Call debug screen 1 to see
VERR
(right column, bottom). TWO (2) possible scenarios
:
1)
IF (and ONLY if) VERR = 0:
For some reason the Valve Error is not occurring at this time.
2)
IF VERR = 1 OR more:
See parts a AND b below:
a) Call debug screen 0.
b) Ignoring the TOP
Flow Error
window, WITHOUT LOOKING
AWAY, watch the
2
nd
window down
,
Valve Error
for one
(1) minute. Ignoring a ‘blink to 1’, that lasts less than one (1)
second, does for
Valve Error
EVER = 1 for LONGER
THAN two (2) seconds?
Yes
Valve Error
= 0 OR ‘blinks’ to 1 for less than one second!
Proceed to
page 195
TROUBLESHOOTING VALVE ERRORS IN DIALYSIS PROGRAM
No
Valve Error
= 1 LONGER THAN two (2) seconds! Proceed to
page 652
..
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...