2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 210
b) Figure below, routing Valve #24’s
INPUT
tubing so that spillage into the hydraulics will
NOT occur, remove it and measure flow from it for thirty (30) seconds.
c) More than 300 ml every 30 seconds?
Yes
More than 300 ml!
A)
Return Valve #24’s tubing;
B)
NOTE
ONLY
VALVE
#24
will be checked and proceed to
page 192
No
Less than 300 ml! Return Valve #24’s tubing THEN see procedure number
(page 210).
VEC- 1.4.1 BAD FLOW TO VALVE #24
a) ENSURE Loading Pressure (gauge in Rinse port) is still PEAKING to between 22 and 27 psi.
b) ENSURE Flow Pump Pressure (gauge at Flow Pump) is still PEAKING to between 34 and 38 psi.
c) If Loading AND Flow Pressure are good, THREE (3) possible bad components:
1)
Actuator-Test Board
OR
2)
Actuator board Cable OR
3)
Multiple bad balancing chamber valves.