2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 314
T- 7.8.0 [MONITOR REFERENCE] CHANGES MORE THAN +/- 0.2° C
a) Figure right
, place the 34° C (6.808 KΩ) plug from the
FOUR RESISTOR SET
into the
2
nd
distribution board
position from the left “MON-NTC”.
b) TWO (2) possible scenarios ,
based on the screen’s
[Monitor
Reference]
data box:
1)
IF (and ONLY if) [Monitor Reference] continues to change more than 0.2 °C per minute:
THREE possible bad components
: 1)
Un12 and/or -12 volt DC voltage (i.e. possible bad
Power Logic Board);
2)
Bad Sensor Board;
3)
Bad Functional Board.
2) IF [Monitor Reference] does NOT change more than 0.2 °C per minute:
NTC #3 may be bad: See
parts a and b below:
a) Replace NTC #3 with a known good! To LOCATE NTC #3 refer to
(page 218).
b) Return to (ABOVE) procedure number
(page 305).
LEFT BLANK INTENTIONALLY