2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 312
T- 7.6.4 ‘SLOW FLOW‘ CALIBRATION
NOTE:
This calibration requires PATIENCE to avoid error!
a)
[TEMP DAC]
= 255 and Flow Rate (Figure right) = 100 (ml/min)! The external flow
indicator’s ‘bob’ may look like it is NOT moving! This is NORMAL here!
b) ‘Sharply’ press ‘CONFIRM’. The screen reads
“11. Adjust the TEMP DAC…”
and
[TEMP DAC]
turns
pale yellow / white.
c) Select
[TEMP DAC].
It turns bright yellow. Set it to “210” then ‘sharply’ press ‘CONFIRM’ ONCE!
d) ENSURING
[TEMP DAC]
is pale yellow / white continue to part e. If (and ONLY if)
[TEMP DAC] is
GRAY escape the calibration and unfortunately return to (ABOVE) procedure number
e) WITHOUT LOOKING AWAY, simultaneously watch the external meter AND the screen’s
[Monitor
Reference]
data box for 13 FULL minutes. TWO (2) possible scenarios:
1)
IF (and ONLY if) one OR both exceed 40.0° C even ONCE:
This time adjusting
[TEMP DAC]
to 10 lower than it already is, repeat parts c through e.
2) IF neither NEVER exceeds 40.0° C:
(page 312).
T- 7.6.5 CALIBRATE SLOW FLOW TEMPERATURE
The external meter fluctuates SLOWLY, approxi/- 0.4° C, between a maximum (high) and a
minimum (low). WITHOUT LOOKING AWAY, watch fifteen (15) consecutive minimum to maximum cycles.
TWO (2) possible scenarios:
1)
IF (and ONLY if) the maximum is consistently between 36.9 and 37.1° C!
See procedure number
(page 313).
2)
IF the maximum is NOT consistently between 36.9 and 37.1° C:
See parts a THROUGH c below:
a) Activate
[TEMP DAC]
, it turns bright yellow and adjust its value (+/- 2 = +/- 0.1° C).
b) ‘Sharply’ press ‘CONFIRM’ to return
[TEMP DAC]
to pale yellow / white. It must NOT turn gray until
the meter consistently fluctuates to a maximum of between 36.9 and 37.1° C.
(page 312).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...