2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 283
T- 3.0.0 IN HEAT DISINFECT TEMPERATURE DID NOT REACH 80° C
a) Without screwing in the rear panels, install the hydraulics into the cabinet!
b)
Place the machine in HEAT DISINFECT!
c) Watch the screen’s
[Temperature]
window for fifteen (15) minutes
OR
until if it increases to 50° C or
more. TWO (2) possible scenarios:
1)
IF (and ONLY if) [Temperature] NEVER reaches at least 50° C:
See procedure number
(page 283).
2) IF [Temperature] INCREASES to 50° C or more:
Continue to watch for
up to twenty (20) FULL
minutes OR until if Heat Disinfect finishes. TWO (2) possible scenarios i) or ii) below:
i)
IF (and ONLY if) [Temperature] DOES NOT reach 80° C OR does but then falls and
REMAINS less than 79° C for at least three (3) minutes:
ii) IF [Temperature] reaches 80° C or more and remains:
The problem (“Temperature does not
reach 80° C) is no longer occurring!
T- 3.0.1 ISOLATE HEATER VOLTAGE
a) Figure right, unplug the connector from the
1st
distribution board position from the left, “CON-NTC”. This
is NTC #2’s connector and unplugging it should turn the
heater on!
b)
Set your CALIBRATED volt meter to AC
voltage (~ V, V
AC
)!
c) Figure right, measure at the distribution board’s Heater
Connector, between the Heater Connector’s BROWN and
BLUE wires.
d) TWO (2) possible scenarios:
1)
IF (and ONLY if) LESS THAN 100.0 volts AC!
Leaving NTC #2 unplugged, proceed to
page 263
2) IF more than 100.0 volts AC:
See parts a AND b below:
a)
Return NTC #2’s connector to the 1
st
distribution board position from the
left “CON-NTC”!
(page 284).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...