2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 440
SECTION 9 - TMP (PRESSURE) PROBLEMS
A) If the drain tubing is transparent, look through its ENTIRE length for possible bio-growth restrictions.
B) Call the Home screen. [
Venous Pressure
] MUST be between -10 and 10 mmHg!
C) The Treatment Clock (Figure right) MUST be off i.e. “Tx Paused”!
D) ENSURE Dialysate Flow has been ON (Flow on/off lamp NOT blinking) for six
(6) minutes.
E) If the Automated Tests are running (screen reads “Test:.…”) allow them to finish!
F)
Remove the ‘dummy venous chamber’ from the Level Detector module till
instructed!
G) Select the [Dialysate Flow] window!
H) Set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’.
I)
DO NOT reset alarms!
J) Is the external flow indicator’s ‘bob’ moving at least ¼ up the sight tube?
Yes
‘Bob’ moving! See procedure number
(page 442).
No
a)
OPEN THE SHUNT DOOR AND LEAVE IT OPEN UNTIL INSTRUCTED!
b) If the TMP window is NOT RED see part K. If red:
A)
Press and
release the ‘Reset’ key then immediately press and hold it for three
(3) seconds;
B)
Allow thirty (30) seconds;
B)
If a TMP alarm
reoccurs attempt RESET up to twice BEFORE continuing to part K!
K) Call debug screen 0. WITHOUT LOOKING AWAY, watch
Flow
Error
for two (2) minutes. If EVER = 1, even just once, indicates
a masked Flow Error! TWO (2) possible scenarios:
1)
IF (and ONLY if) Flow Error ALWAYS = 0!
See part L!
2)
IF Flow Error EVER = 1 even if only once:
Proceed to
page 20
L) Call the Home screen. [
Temperature
] MUST be STABLE
*
between 35.0 and 38.9° C;
[
Conductivity
]
MUST be STABLE
*
between 13.0 and 14.5 mS!
*
STABLE = NOT changing more than 0.2 every minute!
M) At the bottom of the screen, select the ‘Dialysate’ tab!
Parts N through U next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...