2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 619
P- H.1.3 ISOLATE HYDRAULIC SENSORS
a)
IMPORTANT! To prevent damage turn the machine OFF!
b) CAREFULLY
*
return the module cables to their positions.
* CAUTION!
DO NOT a plug a module, other than the Blood Pressure Module, into the Blood
Pressure or Colin position! DO NOT plug the Blood Pressure Module into another module’s
position!
c) CAREFUL HERE!
Per the Figure below
, unplug the SENSOR cable from the Distribution board!
d) Repeat the Voltage Detection Calibration. THREE (3) possible scenarios 1) or 2) or 3):
1)
IF (and ONLY if) “Operator Error” reoccurs:
2) IF (and ONLY if) 5V EST AND / OR 12V EST are not in range:
3)
IF “Operator Error” did NOT occur AND 5V EST AND 12V EST are in range:
One of (12)
twelve Sensors or one of five (5) Pumps may be causing the problem! Perform parts a THROUGH
c below to determine which one:
a)
IMPORTANT!
Turn the machine OFF and reconnect the Sensor Board cable.
b) Per the Figure above, your choice, unplug ONE of 12 hydraulic Sensors or ONE of 5 Pumps
from the Distribution board.
c) Attempt the Voltage Detection Calibration. TWO (2) possible scenarios:
1)
IF “Operator Error” does NOT occur AND 5V EST AND 12V EST are in range:
The
last component unplugged or its cable may be causing the problem.
2) IF problem reoccurs:
Repeat parts b through c until ALL (12) twelve sensors (NTC #2
through NTC #44 AND the Blood Leak Sensor) AND ALL five pumps
(#16 through #22),
have been unplugged.
NOTE:
AFTER all components have been unplugged AND, the
problem reoccurs replace the Sensor Board cable with a known good. If (and ONLY if) the
problem reoccurs see procedure number
(page 620).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...