2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 668
APPENDIX B - CALIBRATE IF REPLACED
Components may need to be replaced to repair the machine. The following table
indicates calibrations that must be performed in this event.
Replaced Component
Then Calibrate
K Calibration
Procedures Section #
- Display Assembly (i.e. screen)
- Touch Screen Calibration
2.1
- Deaeration Pump Head (#20)
- Deaeration and Loading
Pressure
2.2.1
- Deaeration Motor (or if brushes are replaced)
- Loading Pressure Relief Valve (#78)
- Flow Pump Head (#21)
- Flow Pressure
2.2.2
- Flow Motor (or if brushes are replaced)
- Flow Pressure Relief Valve (#78)
- Inlet Water Pressure Regulator (#61)
- Inlet Water Pressure
Regulator
3.1
- Entire Balancing Chamber
- Balance Chamber Volume
2.2.3
- Acid Pump (#16)
- Acid Pump Volume
2.2.4
- Bicarbonate Pump (#17)
- Bicarbonate Pump Volume
2.2.5
- UF Pump (#22)
- UF Pump Volume
2.2.6
- Dialysate Pressure Transducer (#9)
- Dialysate Pressure
2.3.3
- Temperature Sensor (NTC #2 or NTC #3)
- Temperature Control
2.3.6
- Post Dialyzer Temperature Sensor (NTC #44)
- Heater (#54)
- Temperature Control
2.3.6
- Blood Leak Detector (#8)
- Blood Leak
2.3.7
- Pre Dialyzer Conductivity Cell (#7)
- Conductivity Cells
2.3.8
- Post Dialyzer Conductivity Cell (#13)
- Conductivity Cells
2.3.8
- Functional Board (with new EEPROM)
- All Calibrations
All Sections
- Actuator-Test Board
- Voltage Detection
2.4.2
- Sensor Board
- Arterial Pressure
2.3.1
- Venous Pressure
2.3.2
- Dialysate Pressure
2.3.3
- Temperature Sensor
2.3.4
- Post Temperature Sensor
2.3.5
- Temperature Control
2.3.6
- Blood Leak
2.3.7
- Conductivity Cells
2.3.8
- Arterial Pump Rate
2.4.3
- Venous Pump Rate
2.4.4
- Arterial Blood Pump Module
- Arterial Pressure
2.3.1
- Arterial Pump Rate
2.4.3
- Level Detector Module
- Venous Pressure
2.3.2
- Level Detector
3.2
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...